17 Poll the status of the power meter by sending *ESR?. *ESR? will return
In both power sweep and frequency sweep mode, the meter TRIG OUT to power source
N O T E
TRIG IN connection is optional. You may choose to setup the Dwell time in the power
source step setting to cater the maximum settling time required by the power meter.
Determine the Right Step to be Set
Number of frequency step can be calculated by using equation below:
where,
Step = Number of frequency step
f
f
Interval = Frequency step size
Example
When f
Step should be set to
Step = (f
N1913/1914A EPM Series Power Meters User's Guide
a 1 when buffering is completed. Use FETCh? to retrieve all the
buffered measurement.
Step = (f
– f
stop
= Frequency sweep's start point
start
= Frequency sweep's stop point
stop
= 1 GHz and f
start
– f
stop
start
= (5 GHz – 1 GHz + 0.5 GHz)/0.5 GHz
= 9
+ Interval)/Interval
start
= 5 GHz with given interval of 0.5 GHz, the
stop
+ Interval)/Interval
General Power Meter Functions
2
61