General Power Meter Functions
10 Press
11 Set the frequency range and step by sending the below commands to
Refer to the EPM Series Power Meters Programming Guide for the commands usage
N O T E
details.
Example of commands set:
FREQ:STAR 10MHz
FREQ:STOP 500MHz
FREQ:STEP 10
The power meter can be remotely controlled by LAN, USB and GPIB (IEEE488)
N O T E
programming interfaces. Refer to N1913/1914A EPM Series Power Meters Installation
Guide for details.
12 Send *OPC (OPeration Complete) command to meter; to set the
13 Set the meter to continuous trigger cycles by sending
14 Configure and set the required power sweep range and step of the
15 Set the power source trigger input and trigger output.
16 Set the power source to sweep.
60
or
measurement averaging, measurement frequency, offsets, duty cycle and
so forth. Refer to
"General Power Meter Functions"
setup procedures.
meter using remote interface.
• SENSe1:FREQuency:STARt
<start_frequency><frequency_unit>
• SENSe1:FREQuency:STOP <stop_frequency><frequency_unit>
• SENSe1:FREQuency:STEP <frequency_step_size>
operation complete bit in the Standard Event Status register when all
pending device operations are completed.
INITiate:CONTinuous ON command to meter.
power source.
to setup measurement settings such as
N1913/1914A EPM Series Power Meters User's Guide
on page 31 for the