Table 10-5 U2000 Series Power Sensors Specifications - Agilent Technologies N1913a User Manual

Epm series power meters
Hide thumbs Also See for N1913a:
Table of Contents

Advertisement

10
Specifications and Characteristics

Table 10-5 U2000 Series power sensors specifications

The Zero Drift and Measurement Noise specifications are only applicable to U2000 Series
N O T E
USB power sensors with serial prefix as shown below:
U2000A Serial prefix MY480/SG480 and above
U2001A Serial prefix MY481/SG481 and above
U2002A Serial prefix MY482/SG482 and above
U2004A Serial prefix MY484/SG484 and above
For power sensors with earlier prefixes, refer to the U2000 Series Operating and Service
Guide.
226
Power sensor
U2000/1/2A
–60 dBm to –35 dBm
–38 dBm to –15 dBm
–20 dBm to –9 dBm
–11 dBm to –5 dBm
–7 dBm to 15 dBm
10 dBm to 20 dBm
U2004A
–60 dBm to –35 dBm
–38 dBm to –15 dBm
–20 dBm to –9 dBm
–11 dBm to –5 dBm
–7 dBm to 15 dBm
10 dBm to 20 dBm
U2000/1/2H
–50 dBm to –25 dBm
–28 dBm to –5 dBm
–10 dBm to 1 dBm
–1 dBm to 5 dBm
3 dBm to 25 dBm
20 dBm to 30 dBm
U2000/1B
–30 dBm to –5 dBm
–8 dBm to 15 dBm
10 dBm to 21 dBm
19 dBm to 25 dBm
23 dBm to 44 dBm
Range
Zero Drift
Agilent N1913/1914A EPM Series Power Meters User's Guide
1
Measurement Noise
200 pW
400 pW
1.5 nW
1.5 nW
50 nW
650 nW
500 nW
2 µW
200 pW
400 pW
1.5 nW
1.5 nW
50 nW
650 nW
500 nW
2 µW
2 nW
4 nW
15 nW
150 nW
500 nW
6.5 µW
5 µW
20 µW
100 µW
200 nW
400 nW
1.5 µW
1.5 µW
50 nW
650 µW
500 µW
2
1 nW
15 nW
1 µW
10 µW
1 nW
15 nW
1 µW
10 µW
10 nW
15 nW
10 µW
1 µW
15 µW
1 mW

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1914a

Table of Contents