12 Connect the power sensor to the signal to be measured.
13 The corrected measurement result is displayed.

Figure 6-7 Calibration factor displayed

Example
To make a measurement on channel A with a power sensor which has a
reference calibration factor of 99.8% and a calibration factor of 97.6% at
the measurement frequency.
1 Disconnect the power sensor from any signal source.
2 Press
3 Use the numeric keypad to enter 99.8 in the Ref Cal Factor pop- up
4 Press
5 Press
N1913/1914A EPM Series Power Meters User's Guide
%
to complete the entry.
REF CFs
,
window.
%
to complete the entry.
and the channel
displayed.