Example
To make a measurement on channel A with a power sensor which has a
reference calibration factor of 99.8% and a calibration factor of 97.6% at
the measurement frequency.
1 Disconnect the power sensor from any signal source.
2 Press
3 Use the numeric keypad to enter 99.8 in the Ref Cal Factor pop- up
4 Press
5 Press
6 Connect the power sensor to the POWER REF output.
7 Press
8 Press
9 Use the
10 Press
11 Connect the power sensor to the signal to be measured.
12 The corrected measurement result is displayed.
When no sensor tables are selected and
N O T E
calibration factor used for the measurement is displayed in the upper window as shown in
Figure
N1913/1914A EPM Series Power Meters User's Guide
REF CFs
,
window.
%
to complete the entry.
and the channel
displayed.
and the channel
routine. The Calibrating pop- up is then displayed.
Offset
,
and
to display the Cal Factor pop- up. Use the numeric keypad to enter
97.6 in the Cal Factor pop- up window.
%
to complete the entry.
5-5.
Using 8480 Series Power Sensors
and the channel
Zero
softkey. The Zeroing pop- up is
Cal
softkey to start the calibration
. The value is displayed on the Cal Fac field.
keys to highlight the Cal Fac value field and press
Single Numeric
REF CF
softkey.
display mode is chosen, the
5
135