Using E9300 E-Series Power Sensors
5 Press
6 Use the numeric keys to enter the required value.
7 Press
Ensure that the filter is not reset when a step increase or decrease in power is detected by
N O T E
switching the step detection off.
Procedure
Switch off step detection as follows:
1 Press
2 Use the
3 Press
4 Press
Achieving Stable Results with GSM Signals
Signals with a pulse repetition frequency (PRF) close to a multiple or a
submultiple of the 440 Hz chopper- amplifier signal generate a beat note at
a frequency between the PRF and 4F400Hz0 Hz. Control over the filter
settings is again required to obtain stable results.
Tip
114
to display the Meas Avg Count pop- up.
Enter
to complete the entry.
.
On dual channel meters, select the required channel.
and
to check the step detection to Off.
key to close the Channel Setup screen.
The PRF of a GSM signal is approximately 217 Hz and thus requires
more averaging than most other TDMA signals. To achieve a stable
measurement use the filter setting procedures to set the
Length
Experimentally, a
although settings in the order of 31 or 32 give acceptable results if a
faster measurement is required.
keys to select the Step Detect setting field.
setting of 148 gives optimum results
N1913/1914A EPM Series Power Meters User's Guide