Setting External Trigger For Average Power Measurement - Agilent Technologies N1913a User Manual

Epm series power meters
Hide thumbs Also See for N1913a:
Table of Contents

Advertisement

2
General Power Meter Functions

Setting External Trigger for Average Power Measurement

There are two modes featured in triggered average power measurement
• Power sweep mode
• Frequency sweep mode
These modes are used to eliminate the need for lengthy test routines,
while increasing measurement throughput by reducing overhead of
communication with the controller. The sweep feature allows you to make
power measurement by quickly stepping through a series of frequencies or
power levels. Configuration shown in
connection required to synchronize power meter measurement to the
power source settings.
Trigger delay and trigger holdoff are both not applicable when the power meter is set to
N O T E
power sweep mode or frequency sweep mode.
54
Figure 2- 19
N1913/1914A EPM Series Power Meters User's Guide
illustrates the triggering

Hide quick links: