Frequency Sweep Mode
Frequency sweep is generally used in a frequency response calibration
system where the amplitude is fixed, and the frequency of the power
source signal is swept. This mode can be used to determine the frequency
response of a device under test.
This feature is only available when N8480 Series, E-Series E4410, or E-Series E9300 sensor
N O T E
is connected.
Procedures
1 Connect sensor to a power source.
2 Connect power meter TRIG OUT to power source TRIG IN using a
3 Press
When N8480 Series, E-Series E4410, E-Series E9300 or U2000 Series sensor is connected,
N O T E
Sensor Mode is set to AVG only by default.
4 Press
5 Press
6 Select either
7 Press
8 Press
9 Press
N1913/1914A EPM Series Power Meters User's Guide
BNC cable. The same connection is applied for power source TRIG OUT
to power meter TRIG IN. See
. The Channel Setup displays as shown in
. The Trigger menu is displayed.
Acqn
softkey to configure a trigger.
Sing Trig
Settings
to configure the remaining trigger parameters. The trigger
Settings menu consists of two pages.
Figure 2- 22
shows page 2.
Source
Ext
and
1 of 2
Output
,
at the rear panel TRIG OUT BNC port when the meter is triggered upon a
completed measurement.
Figure 2-
19.
Cont Trig
or
.
Figure 2- 21
will be automatically enabled.
On
and
to allow a TTL level high to be produced
General Power Meter Functions
Figure
2-20.
shows page 1 and
2
59