Using E4410 E-Series Power Sensors; Figure 4-3 Frequency Pop-Up - Agilent Technologies N1913a User Manual

Epm series power meters
Hide thumbs Also See for N1913a:
Table of Contents

Advertisement

4

Using E4410 E-Series Power Sensors

You can reduce the steps required to carry out the zero and calibration procedure as
N O T E
follows:
1 Connect the power sensor to the POWER REF output.
2 Press
Now, set the frequency of the signal you want to measure. The power
meter automatically selects the appropriate calibration factor.
6 Press
7 Use the

Figure 4-3 Frequency pop-up

8 Confirm your choice by pressing
9 Press
10 Connect the power sensor to the signal to be measured.
124
Zero + Cal
and
Zero + Cal
Zero + Cal A
,
. On dual channel meters select the required channel.
and
press
to display the Frequency pop- up. Use the numeric keypad to
enter the required value in the Frequency pop- up window.
key to close the Channel Setup screen.
Make the measurement now.
The corrected measurement result is displayed.
. (For dual channel meters, press
Zero + Cal B
or
as required).
keys to highlight the Frequency value field and
MHz
or
N1913/1914A EPM Series Power Meters User's Guide
GHz
.