Figure 6-6 Calibration Factor Pop-Up Window - Agilent Technologies N1913a User Manual

Epm series power meters
Hide thumbs Also See for N1913a:
Table of Contents

Advertisement

6
Using N8480 Series Power Sensors
You can reduce the steps required to carry out the zero and calibration procedure as
N O T E
follows:
1 Connect the power sensor to the POWER REF output.
2 Press
Now, set the sensor calibration factor for the frequency of the signal to
measure.
9 Check the current reference calibration factor setting by pressing
Does this setting match the value for the sensor?
N O T E
The power sensor reference calibration factor can normally be found above the calibration
factors table on the power sensor body.
10 To change the settings, use the

Figure 6-6 Calibration factor pop-up window

156
Zero + Cal
and
Zero + Cal
Zero + Cal A
,
Offset
. The value is displayed on the Cal Fac field.
,
value field and press
numeric keypad to enter the required value in the Cal Factor pop- up
window.
. (For dual channel meters, press
Zero + Cal B
or
as required).
and
to display the Cal Factor pop- up. Use the
N1913/1914A EPM Series Power Meters User's Guide
keys to highlight the Cal Fac