2
General Power Meter Functions
Refer to the N1913/1914A EPM Series Power Meters Programming Guide for the
N O T E
commands usage details.
Example of command set:
BUFF:COUN 100
The power meter can be remotely controlled by LAN, USB and GPIB (IEEE488)
N O T E
programming interfaces. Refer to N1913/1914A EPM Series Power Meters Installation
Guide for details.
12 Press
13 Send *OPC (OPeration Complete) command to meter to set the
14 Set the meter to continuous trigger mode by sending
15 Configure and set the required power sweep range and step on the
16 Set the power source trigger input and trigger output, then start
17 Poll the status of the power meter by sending *ESR?. *ESR? will return
58
or
measurement averaging, measurement frequency, offsets, duty cycle and
so forth. Refer to
"General Power Meter Functions"
setup procedures.
operation complete bit in the Standard Event Status register when all
pending device operations are completed.
INITiate:CONTinuous ON command to meter.
power source accordingly.
sweeping.
a 1 when buffering is completed. Use FETCh? to retrieve all the
buffered measurement.
to setup measurement setting such as
N1913/1914A EPM Series Power Meters User's Guide
on page 31 for the