Resolution is a measurement display function and not a channel function.
In the case where a channel is set up in both the upper and lower
window and the resolution settings are different, the highest resolution
setting is taken to calculate the averaging number.
These four resolution levels represent:
• 1, 0.1, 0.01, 0.001 dB respectively if the measurement suffix is dBm or
• 1, 2, 3 or 4 significant digits respectively if the measurement suffix is
Procedure
Set measurement averaging as follows:
1 Press
2 Use the
3 Press
4 Use the
5 Press

Figure 2-15 Meas Avg Count pop-up

6 Use the numeric keys to enter the required value and press
7 Press
50
dB.
Ω or %.
. On dual channel meters select the required channel. The
current setting is shown in the Meas Avg field (
the Channel Setup screen. The default is
and
and use the
settings. If you have selected AUTO or OFF proceed at step 7. If you have
selected MAN proceed as follows.
key to select the Meas Avg: value field.
to display the Meas Avg Count pop- up.
key to close the Channel Setup screen.
AUTO
keys to select the Filter setting field.
and
to step through the available
N1913/1914A EPM Series Power Meters User's Guide
, MAN, or
AUTO
.
Enter
) on
OFF
.