N1913/1914A EPM Series Power Meters User's Guide
Sensor Model
Agilent 8485D
Agilent 8487A
Agilent 8487D
Connection Requirements
Prior to the power meter being zeroed and calibrated, an Agilent
11708A 30 dB reference attenuator and an APC 3.5 (f) to 50 Ω (m)
N-Type adapter (08485-60005) should be connected between the
power sensor and the reference calibrator. Remove this attenuator
from the power sensor input before making measurements.
This power sensor requires an APC 2.4 (f) to 50 Ω (m) N-Type adapter
(08487-60001) to connect to the power meter. Remove this attenuator
before making measurements.
Prior to the power meter being zeroed and calibrated, an Agilent
11708A 30 dB reference attenuator and an APC 2.4 (f) to 50 Ω (m)
N-Type adapter (08487-60001) should be connected between the
power sensor and the reference calibrator. Remove this attenuator
from the power sensor input before making measurements.
General Power Meter Functions
2
83