Table 10-2 Zero Set (Internal And External) For U2000 Series - Agilent Technologies N1913a User Manual

Epm series power meters
Hide thumbs Also See for N1913a:
Table of Contents

Advertisement

10
Specifications and Characteristics

Table 10-2 Zero Set (Internal and External) for U2000 Series

The Zero Set specifications are only applicable to U2000 Series USB power sensors with
N O T E
serial prefix as shown below:
U2000A Serial prefix MY480/SG480 and above
U2001A Serial prefix MY481/SG481 and above
U2002A Serial prefix MY482/SG482 and above
U2004A Serial prefix MY484/SG484 and above
For power sensors with earlier prefixes, refer to the Agilent U2000 Series USB Power
Sensors Operating and Service Guide.
222
Power sensor
U2000/1/2A
–60 dBm to –35 dBm
–38 dBm to –15 dBm
–20 dBm to –9 dBm
–11 dBm to –5 dBm
–7 dBm to 15 dBm
10 dBm to 20 dBm
U2004A
–60 dBm to –35 dBm
–38 dBm to –15 dBm
–20 dBm to –9 dBm
–11 dBm to –5 dBm
–7 dBm to 15 dBm
10 dBm to 20 dBm
U2000/1/2H
–50 dBm to –25 dBm
–28 dBm to –5 dBm
–10 dBm to 1 dBm
–1 dBm to 5 dBm
3 dBm to 25 dBm
20 dBm to 30 dBm
U2000/1B
–30 dBm to –5 dBm
–8 dBm to 15 dBm
10 dBm to 21 dBm
19 dBm to 25 dBm
23 dBm to 44 dBm
Range
Zero Set (Internal)
Agilent N1913/1914A EPM Series Power Meters User's Guide
Zero Set (External)
±1.5 nW
±600 pW
±2 nW
±1.5 nW
±12 nW
±10 nW
±2 µW
±500 nW
±4 µW
±1 µW
±6 µW
±5 µW
±2.8 nW
±600 pW
±3 nW
±1.5 nW
±12 nW
±10 nW
±2 µW
±500 nW
±4 µW
±1 µW
±6 µW
±5 µW
±15 nW
±8 nW
±20 nW
±20 nW
±120 nW
±100 nW
±20 µW
±20 µW
±40 µW
±30 µW
±60 µW
±60 µW
±1.8 µW
±800 nW
±2 µW
±2 µW
±12 µW
±10 µW
±2 mW
±1 mW
±4 mW
±2 mW