Figure 2-9 Frequency Dependent Offset Indicator - Agilent Technologies N1913a User Manual

Epm series power meters
Hide thumbs Also See for N1913a:
Table of Contents

Advertisement

2 Use the
When no data is contained in the highlighted table, the
N O T E
3 Press
4 Press
Offset Table A Selected

Figure 2-9 Frequency dependent offset indicator

5 To change the frequency, press
6 Press
7 To confirm your choice, press the appropriate unit softkey.
8 Connect the power sensor to the signal to be measured.
9 The measurement result, including offset, is now displayed.
If the measurement frequency does not correspond directly to a frequency in the sensor
N O T E
calibration table (if selected) and the frequency dependent offset table being used, the
power meter calculates the calibration factor and offset using linear interpolation.
If you enter a frequency outside the frequency range defined in the sensor calibration table
or the frequency dependent offset table, the power meter uses the highest or lowest
frequency point in the appropriate table to set the calibration factor and offset.
N1913/1914A EPM Series Power Meters User's Guide
and
Table
A Table
press
(
to complete the selection of the offset table.
again to display the measurement screen.
shows which offset table is selected.
to highlight the Frequency field.
to display the Frequency pop- up window. Use the numeric
keypad to enter the required value in the Frequency pop- up window.
General Power Meter Functions
keys to highlight one of the 10 table titles and
B Table
or
for dual channel) to highlight
and use the
Table
key is disabled (grayed out).
Figure 2- 9
and
2
On
.
keys
43