Figure 2-19 Trig In And Trig Out Connection Diagram Between Power Meter And - Agilent Technologies N1913a User Manual

Epm series power meters
Hide thumbs Also See for N1913a:
Table of Contents

Advertisement

Power Sweep Mode
Power sweep is generally used in power level calibration setup where the
frequency is fixed (CW frequency), and the amplitude of the power source
signal is swept. This mode can be used to characterize the flatness,
linearity or gain compression of a device under test.
This feature is only available when 8480 Series, N8480 Series, E-Series E4410, or E-Series
N O T E
E9300 sensor is connected.
Procedures
1 Connect sensor to a power source.
2 Connect power meter TRIG OUT to power source TRIG IN using a
Figure 2-19 TRIG IN and TRIG OUT connection diagram between power meter and power
3 Press
N1913/1914A EPM Series Power Meters User's Guide
BNC cable. The same connection is applied for power source TRIG OUT
to power meter TRIG IN.
TRIG
TRIG
IN
OUT
Power Source
source
. The Channel Setup screen displays as below.
General Power Meter Functions
(optional connection)
Average Power Sensor
TRIG
TRIG
OUT
IN
Power Meter
2
55

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

N1914a

Table of Contents