Electromagnetic Compatibility (Emc) Measurements - Agilent Technologies N1913a User Manual

Epm series power meters
Hide thumbs Also See for N1913a:
Table of Contents

Advertisement

7
Using U2000 Series USB Power Sensors

Electromagnetic Compatibility (EMC) Measurements

The low frequency range of the U2004A make it the ideal choice for
making EMC measurements to CISPR (Comite International Special
Perturbations Radioelectriques) requirements, and electromagnetic
interference (EMI) test applications such as the radiated immunity test
(IEC61000- 4- 3).
DC coupling of the U2004A input allows excellent low frequency coverage.
However, the presence of any dc voltages mixed with the signal has an
adverse effect on the accuracy of the power measurement.
The U2004A sensor is DC coupled. DC voltages in excess of the maximum value (5 VDC)
C A U T I O N
can damage the sensing diode.