Status Messages, Troubleshooting; Blinking Screen - ABB EL6010 Operator's Manual

Continuous gas analyzers
Table of Contents

Advertisement

Chapter 8

Blinking Screen

Blinking Screen
Measured value
exceeds / is below
measuring range span
Maintenance
Request
Failure
Error List
42/24-30 EN Rev. 1

Status Messages, Troubleshooting

Blinking of part of the screen in measuring mode indicates that the gas analyzer or
process system has experienced a fault.
CO
753.2
ppm
CO
105.2 ppm
CO
105.2
ppm
Error list
30
Error list
no errors
EL6010 Gas Analyzer in Category 2G Operator's Manual
The measured value reading blinks if the measured
value of the sample component is < –100 % or
130 % of measuring range span.
>
The sample component name display blinks if a
maintenance request condition has occurred.
The gas analyzer is in a state that will soon require
user intervention. The measured value is valid.
The sample component name and measured value
displays blink if a failure condition has occurred.
The gas analyzer is in a state that immediately
requires user intervention. The measured value is
invalid.
CO
In the event of faults, the status message numbers
are displayed in the Error List menu item for
each sample component (see page 46).
To select the menu item, press the
CO
This display indicates there are no errors present.
key twice.
45

Advertisement

Table of Contents
loading

Table of Contents