78
Status messages
Information displayed on the ship control system
Information displayed on the analyzer system
Sample gas cooler problems
Problem
No display
Cooler not running
Status-LED blinks with high
temperature
Status-LED blinks with low
temperature
Condensate in gas outlet
Reduced gas flow
Display toggles between
temperature and error
message Error 1
Display toggles between
temperature and error
message Error 2
GAA330-M EMISSION MONITORING SYSTEM
•
Calibration analyzer
Back-flush active
•
Collective error
•
Measuring range selection SO
•
MR1 0 to 250 ppm
•
MR2 0 to 500 ppm
•
Temperature alarm
•
Fault cooler
•
Fault humidity
•
Flow alarm min.
•
Flow alarm max.
•
Condensate level
•
Cause
No power
Fuse blown
High temperature at the
compressor casing
Operational temperature
not yet reached
Cooling capacity to low, even
though cooler is running
Gas flow / dew point / gas
temperature too high
Fan broken
Control defect
Condensate flask full
Stuck valve in automatic
condensate drain
Cooler overloaded
Clogged gas path
Condensate outlet clogged
by ice
Broken wire
Short circuit
|
OI/GAA330-M-EN REV. C
2
Solution
Check power supply
Check fuse and change it if necessary
Wait until cooled off and care for enough ven-
tilation
Wait for 20 minutes maximum
Make sure that air can circulate free and that
ventilation louvers are not obstructed
Check application parameters, install
pre-separator
Check fan, replace if necessary
Send cooler for inspection
Drain flask
Flush both directions
Check limiting parameters
Check / flush heat exchanger
Send cooler for inspection
Temperature sensor defect. Send cooler for
repair
Temperature sensor defect. Send cooler for
repair