Status Messages - ABB ACF-NT Operator's Manual

Multicomponent analyzer systems for emission and process monitoring
Hide thumbs Also See for ACF-NT:
Table of Contents

Advertisement

Status Messages

No.
Message Text
Syst. Failure
Syst. Maint. Req.
Syst. Maintenance
Module Failure
FID Error
FID Maint. req.
FID Maint. mode
O2 Error
O2 Maint. req.
O2 Maint. mode
110 Sys. Boot
1001 Pump ctrl.
10000 Emerg. Purge
10001 Message back purge
10002 FTIR Failure
10003 FTIR Maint. Req.
10004 FTIR Maintenance
10005 Temp MGL
10005 Temp Pipe
10005 Temp Probe
10007 Flow spectrom.
10008 Service
10009 Enablir failure
10010 Enablir Maint. Req.
10011 Enablir Maintenance
10013 Com watchdog
10014 Sequence trigger
10015 Zero Y04
10016 Span Y05
10017 Error save data
10018 Shut down
10019 High conc.
10020 Purge locked
10021 Start Purge
10022 ZRO2
38
F
MR
MM Comments
X
Measured value is invalid.
X
Measured value is valid.
X
Discard the measured value as a process measured value.
Error on Module RGM11, MultiFID14, FTIR.
FID measured value is invalid.
X
The analyzer is in a state that will soon require user
intervention.
X
The analyzer is being calibrated or serviced.
O2 measured value is invalid.
X
The analyzer is in a state that will soon require user
intervention.
X
The analyzer is being calibrated or serviced.
X
Emergency purge, check temp error (FTIR, probe, pipe,
sample gas line).
X
X
Contact service.
X
Contact service.
X
X
Temp. error sample gas line.
X
Temp. error pipe.
X
Temp. error sample probe.
X
Flow to FTIR is too low.
X
Key "Service" is activated.
X
Contact service.
X
X
X
Contact service.
X
Key "Shut down" is activated.
X
Zero gas valve is activated.
X
Span gas valve is activated.
Shut down process is running.
X
Concentration too high.
X
Key "Start purging" is activated.
X
Detector measuring value exceeds the –10000 to +5000
range. Oxygen-free test gas is being used.
F
= Failure
MR = Maintenance Request
MM = Maintenance Mode
Operator's Manual ACF-NT, ACF-NT V0309
42/23-572 EN Rev. 5

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Acf-nt v0309

Table of Contents