Section 11: Calibration
AC current calibration
Make the connections as shown, then perform the following calibration steps (AC Cal Step 11
through Step 16):
Figure 11-6: AC current verification 1mA-1A ranges
AC Cal Step 11: 100uA @ 1kHz
1. Send the following commands:
SEND dmm.func = dmm.AC_CURRENT
SEND dmm.range = 100e-6
2. Source 100µA @ 1kHz
3. Send the following command:
SEND dmm.calibration.ac(11) print("done")
AC Cal Step 12: 1mA @ 1kHz
1. Send the following command:
SEND dmm.range = 1e-3
2. Source 1mA @ 1kHz
3. Send the following command:
SEND dmm.calibration.ac(12) print("done")
11-14
Series 3700 System Switch/Multimeter Reference Manual
Document Number: 3700S-901-01 Rev. A / August 2007