Section 8: Status Model
The used bits of the System Event Registers are described as follows:
System summary event commands appear in the following table:
Commands
To set register bits:
status.system.enable = status.system.EXTENSION_BIT
status.system.enable = status.system.EXT
status.system.enable = status.system.NODEn
To read registers:
print(status.system.enable)
print(status.system.condition)
print(status.system.event)
To set register bits:
status.system2.enable = status.system.EXTENSION_BIT
status.system2.enable = status.system.EXT
status.system2.enable = status.system.NODEn
To read registers:
print(status.system2.enable)
print(status.system2.condition)
print(status.system2.event)
To set register bits:
status.system3.enable = status.system.EXTENSION_BIT
status.system3.enable = status.system.EXT
status.system3.enable = status.system.NODEn
To read registers:
print(status.system3.enable)
print(status.system3.condition)
print(status.system3.event)
To set register bits:
status.system4.enable = status.system.EXTENSION_BIT
status.system4.enable = status.system.EXT
status.system4.enable = status.system.NODEn
To read registers:
print(status.system4.enable)
print(status.system4.condition)
print(status.system4.event)
8-14
EXT: Set bit indicates that an extension bit from a another system status register is set.
NODEn: Indicates a bit on TSP-Link node n has been set (n =1 to 64).
Series 3700 System Switch/Multimeter Reference Manual
Document Number: 3700S-901-01 Rev. A / August 2007
Bit
B0
B0
Bn
B0
B0
Bn
B0
B0
Bn
B0
B0
Bn