Keithley 3700 series Reference Manual page 339

System switch/multimeter
Hide thumbs Also See for 3700 series:
Table of Contents

Advertisement

Series 3700 System Switch/Multimeter Reference Manual
status.operation.user.*
Remarks
Example
Document Number: 3700S-901-01 Rev. A / August 2007
operreg can also be set to the decimal weight of the bit to be set:
To set user BIT0, set operreg to 1 (20).
To set user BIT4, set operreg to 16 (24). To set user BIT11, set operreg to 2048
(211).
To set more than one bit of the register, set operreg to the sum of their decimal
weights. For example, to set BIT0 and BIT4, set operreg to 17 (1 + 16).
This attribute is used to read or write to the operation user registers.
Bits of the user event register are set by setting the corresponding bits of the user
enable register and the user condition register. For example, the following code
will set B1 (Bit 1) of the user event register:
status.operation.user.enable = 2
status.operation.user.condition = 2
Reading a status register returns a value. The binary equivalent of the returned
value indicates which register bits are set. The least significant bit of the binary
number is bit 0, and the most significant bit is bit 15.
For example, assume value 17 is returned for the enable register. The binary
equivalent is 0000000000010001. This value indicates that BIT0 and BIT4 are set.
Sets user BIT0 of the operation user enable register:
status.operation.user.enable =
status.operation.user.BIT0
Section 9: Instrument Control Library (ICL)
.condition
.enable
.event
.ntr
.ptr
9-195