Section 10: Verification
Applied current
Verify AC curr 1mA @ 1kHz
Verify AC curr 10mA @ 1kHz
Verify AC curr 100mA @ 1kHz 1.00E-01
Verify AC curr 1A @ 1kHz
AC current verification data 3A range
Use the following values to verify the performance of the Series 3700. Actual values are
dependent on published specifications. See
Applied current
Verify AC curr 3A @ 1kHz

Verifying frequency

To verify the Series 3700 frequency function:
1. Connect the Agilent 33220A function generator to the Series 3700 INPUT pins.
2. Set the function generator to output a 1kHz, 5V RMS sine wave.
3. Select the Series 3700 frequency function by pressing the FREQ key.
4. Verify that the Series 3700 each frequency reading is within the limits contained in the table
contained in Frequency verification data.
10-14
Series 3700 System Switch/Multimeter Reference Manual
Range
Current
1.00E-03
1.00E-03
1.00E-02
1.00E-02
1.00E-01
1.00E+00
1.00E+00
Example reading limit calculation
Range
Current
3.00E+00
3.00E+00
Document Number: 3700S-901-01 Rev. A / August 2007
Lower limit
Upper limit
9.989000E-04
1.001100E-03
9.989000E-03
1.001100E-02
9.989000E-02
1.001100E-01
9.977000E-01
1.002300E+00
Lower limit
Upper limit
2.993100E+00
3.006900E+00
(on page 10-3).