Dc Current Calibration; Figure 11-4: Dc Current Verification 10Ua-100Ua Ranges - Keithley 3700 series Reference Manual

System switch/multimeter
Hide thumbs Also See for 3700 series:
Table of Contents

Advertisement

Series 3700 System Switch/Multimeter Reference Manual

DC current calibration

Make the connections as shown, then perform the following calibration steps (DC Cal Step 10
through Step 14):

Figure 11-4: DC current verification 10uA-100uA ranges

DC Cal Step 10: 100uA
1. Send the following commands:
SEND dmm.func = dmm.DC_CURRENT
SEND dmm.range = 100e-6
2. Source 100µA.
3. Send the following commands:
SEND dmm.calibration.dc(10,.0001) print("done")
Document Number: 3700S-901-01 Rev. A / August 2007
Section 11: Calibration
11-9