Section 10: Verification
Figure 10-3: DC current verification 10uA-100uA ranges
DC current verification data 10uA-100uA ranges
Use the following values to verify the performance of the Series 3700. Actual values are
dependent on published specifications. See
Applied current
Verify DC curr 10A
Verify DC curr 10A
Verify DC curr 10A
Zero 100A
Verify DC curr 100A
Verify DC curr 100A
10-10
Series 3700 System Switch/Multimeter Reference Manual
Example reading limit calculation
Range
Current
1.00E-05
0.00E+00
1.00E-05
1.00E-05
1.00E-05
-1.00E-05
1.00E-04
1.00E-04
1.00E-04
-1.00E-04
Document Number: 3700S-901-01 Rev. A / August 2007
Lower limit
Upper limit
-3.000000E-10
3.000000E-10
9.996700E-06
1.000330E-05
-1.000330E-05
-9.996700E-06
9.996910E-05
1.000309E-04
-1.000309E-04
-9.996910E-05
(on page 10-3).