Figure 10-3: Dc Current Verification 10Ua-100Ua Ranges - Keithley 3700 series Reference Manual

System switch/multimeter
Hide thumbs Also See for 3700 series:
Table of Contents

Advertisement

Section 10: Verification

Figure 10-3: DC current verification 10uA-100uA ranges

DC current verification data 10uA-100uA ranges
Use the following values to verify the performance of the Series 3700. Actual values are
dependent on published specifications. See
Applied current
Verify DC curr 10A
Verify DC curr 10A
Verify DC curr 10A
Zero 100A
Verify DC curr 100A
Verify DC curr 100A
10-10
Series 3700 System Switch/Multimeter Reference Manual
Example reading limit calculation
Range
Current
1.00E-05
0.00E+00
1.00E-05
1.00E-05
1.00E-05
-1.00E-05
1.00E-04
1.00E-04
1.00E-04
-1.00E-04
Document Number: 3700S-901-01 Rev. A / August 2007
Lower limit
Upper limit
-3.000000E-10
3.000000E-10
9.996700E-06
1.000330E-05
-1.000330E-05
-9.996700E-06
9.996910E-05
1.000309E-04
-1.000309E-04
-9.996910E-05
(on page 10-3).