Keithley 3700 series Reference Manual page 342

System switch/multimeter
Hide thumbs Also See for 3700 series:
Table of Contents

Advertisement

Section 9: Instrument Control Library (ICL)
status.request_enable
Attribute
Usage
9-198
Service request enable register.
Reads service request enable register:
servenabreg = status.request_enable
Writes to system enable register:
status. request_enable = servenabreg
Set servenabreg to one of the following values:
0 Clears all bits.
To set (enables) MSB bit (B0):
status.MEASUREMENT_SUMMARY_BIT
- or - status.MSB
To set (enables) SSB bit (B1): status.SYSTEM_SUMMARY_BIT
- or - status.SSB
To set (enables) EAV bit (B2): status.ERROR_AVAILABLE
- or - status.EAV
To set (enables) QSB bit (B3):
status.QUESTIONABLE_SUMMARY_BIT
- or - status.QSB
To set (enables) MAV bit (B4): status.MESSAGE_AVAILABLE
- or - status.MAV
To set (enables) ESB bit (B5): status.EVENT_SUMMARY_BIT
- or - status.ESB
To set (enables) OSB bit (B7): status.OPERATION_SUMMARY_BIT
- or - status.OSB
servenabreg can also be set to the decimal weight of the bit to be set:
To set bit B0 (MSB), set servenabreg to 1 (20).
To set bit B2 (EAV), set servenabreg to 4 (22).
To set bit B7 (OSB), set servenabreg to 128 (27).
To set more than one bit of the register, set servenabreg to the sum of their decimal
weights. For example, to set bits B0 and B7, set servenabreg to 129 (1 + 128).
Series 3700 System Switch/Multimeter Reference Manual
Document Number: 3700S-901-01 Rev. A / August 2007

Advertisement

Table of Contents
loading

Table of Contents