Series 3700 System Switch/Multimeter Reference Manual
scan.trigger.clear
Function
Usage
Remarks
Example
scan.trigger.measure.clear
Function
Usage
Remarks
Example
scan.trigger.measure.set
Function
Usage
Remarks
Example
scan.trigger.measure.stimulus
Attribute
Usage
Document Number: 3700S-901-01 Rev. A / August 2007
Clear the trigger model.
scan.trigger.clear()
This function will set the channel, measure and sample event detectors of the trigger
model to the undetected state.
To clear the trigger model:
scan.trigger.clear()
Clear the measure event detector.
scan.trigger.measure.clear()
This function will set the measure event detector to the undetected state.
To clear the measure event detector:
scan.trigger.measure.clear()
Set the measure event detector to the detected state.
scan.trigger.measure.set()
This function will set the measure event detector to the detected state.
To set the measure event detector to the detected state:
scan.trigger.measure.set()
Measure event detector trigger stimulus selection.
To read the trigger stimulus:
eventid = scan.trigger.measure.stimulus
eventid: the present trigger stimulus being used for the measure event
To write the trigger stimulus:
scan.trigger.measure.stimulus = eventid
eventid: the trigger stimulus to set for the measure event
Section 9: Instrument Control Library (ICL)
9-171