7. Install the Fluke 5724A amplifier.
8. Source AC current for the 3A range test points summarized in the AC current verification
data table. Be sure that the 3A readings are within stated limits.

Figure 10-5: AC current verification 1mA-1A ranges

Figure 10-6: AC current verification 3A range

AC current verification data 1mA-1A ranges
Use the following values to verify the performance of the Series 3700. Actual values are
dependent on published specifications. See
Document Number: 3700S-901-01 Rev. A / August 2007
Example reading limit calculation
Section 10: Verification
(on page 10-3).