Figure 10-1: Dc Volts Verification - Keithley 3700 series Reference Manual

System switch/multimeter
Hide thumbs Also See for 3700 series:
Table of Contents

Advertisement

Section 10: Verification
6. Source positive and negative full-scale and half-scale voltages for each of the ranges listed
below. For each voltage setting, be sure that the reading is within stated limits.

Figure 10-1: DC volts verification

DCV verification data
Use the following values to verify the performance of the Series 3700. Actual values are
dependent on published specifications (see
Applied voltage
REL U.U.T. for 100mV
Verify DCV 100mV
Verify DCV 100mV
Verify DCV 100mV
Verify DCV 100mV
REL U.U.T. again for 1V
Verify DCV 1V
Verify DCV 1V
Verify DCV 1V
Verify DCV 1V
Disable REL
Verify DCV 10V
Verify DCV 10V
Verify DCV 10V
Verify DCV 10V
Verify DCV 10V
Disable REL
10-6
Series 3700 System Switch/Multimeter Reference Manual
Example reading limit calculation
Range
Voltage
1.00E-01
0V
1.00E-01
1.00E-01
1.00E-01
5.00E-02
1.00E-01
-5.00E-02
1.00E-01
-1.00E-01
1.00E+00
0V
1.00E+00
1.00E+00
1.00E+00
5.00E-01
1.00E+00
-5.00E-01
1.00E+00
-1.00E+00
1.00E+01
1.00E+01
1.00E+01
5.00E+00
1.00E+01
0.00E+00
1.00E+01
-5.00E+00
1.00E+01
-1.00E+01
Document Number: 3700S-901-01 Rev. A / August 2007
Lower limit
Upper limit
9.999610E-02
1.000039E-01
4.999760E-02
5.000240E-02
-5.000240E-02
-4.999760E-02
-1.000039E-01
-9.999610E-02
9.999680E-01
1.000032E+00
4.999830E-01
5.000170E-01
-5.000170E-01
-4.999830E-01
-1.000032E+00
-9.999680E-01
9.999730E+00
1.000027E+01
4.999855E+00
5.000145E+00
-2.000000E-05
2.000000E-05
-5.000145E+00
-4.999855E+00
-1.000027E+01
-9.999730E+00
(on page 10-3)).

Advertisement

Table of Contents
loading

Table of Contents