Keithley 3700 series Reference Manual page 314

System switch/multimeter
Hide thumbs Also See for 3700 series:
Table of Contents

Advertisement

Section 9: Instrument Control Library (ICL)
scan.trigger.channel.set
Function
Usage
Remarks
Example
scan.trigger.channel.stimulus
Attribute
Usage
Remarks
Example
9-170
Set the channel event detector to the detected state.
scan.trigger.channel.set()
This function will set the channel event detector to the detected state.
scan.trigger.channel.set()
Channel event detector stimulus selection.
To read the trigger stimulus:
eventid = scan.trigger.channel.stimulus
eventid: Present trigger stimulus being used for the channel action.
To write the stimulus:
scan.trigger.channel.stimulus = eventid
eventid: Trigger stimulus to set for the channel action
This attribute selects which event(s) will cause the channel event detector to enter the
detected state. Set this attribute to nil to bypass waiting for an event.
eventid may be one of the following (existing trigger event ID's):
digio.trigger[line].EVENT_ID: An edge (either rising, falling, or either based on the
configuration of the line) on the digital input line.
display.trigger.EVENT_ID: The trigger key on the front panel is pressed.
trigger.EVENT_ID: A *trg message on the active command interface. If GPIB is
the active command interface, a GET message will also generate this event. Also,
the trigger.set function will generate this event.
trigger.blender[N].EVENT_ID: A combination of events has occurred.
trigger.timer[N].EVENT_ID: A delay expired. These can be expanded to include
LXI alarms.
tsplink.trigger[N].EVENT_ID: An edge (either rising, falling, or either based on the
configuration of the line) on the tsplink trigger line.
lan.trigger[N].EVENT_ID
scan.trigger.EVENT_SCAN_READY: Scan Ready Event.
scan.trigger.EVENT_SCAN_START: Scan Start Event
scan.trigger.EVENT_CHANNEL_READY: Channel Ready Event
scan.trigger.EVENT_MEASURE_COMP: Measure Complete Event
scan.trigger.EVENT_SEQUENCE_COMP: Sequence Complete Event
scan.trigger.EVENT_SCAN_COMP: Scan Complete Event
scan.trigger.EVENT_IDLE: Idle Event
To set trigger stimulus of the channel event detector to armed event:
scan.trigger.channel.stimulus =
scan.trigger.EVENT_ARMED
Series 3700 System Switch/Multimeter Reference Manual