Series 3700 System Switch/Multimeter Reference Manual
2. Wait during the five minute delay.
3. Perform the following calibration steps (DC Cal Step 0 through Step 5):

Figure 11-1: Four-wire short diagram

DC Cal Step 0: A/D Mux Offset Cal
Send the following commands:
SEND dmm.calibration.dc(0) print("done")
SEND print(errorqueue.count)
DC Cal Step 1: Input Short Circuit
1. Wait during the 30 second delay.
2. Send the following commands:
SEND dmm.calibration.dc(1) print("done")
SEND print(errorqueue.count)
NOTE
Document Number: 3700S-901-01 Rev. A / August 2007
DO NOT install cables to the inputs (cables will be installed in DC Cal Step 3).
'Check for errors after each step
Section 11: Calibration
11-5