Verifying Two-Wire Resistance; Figure 10-7: Verifying Four-Wire Resistance - Keithley 3700 series Reference Manual

System switch/multimeter
Hide thumbs Also See for 3700 series:
Table of Contents

Advertisement

Section 10: Verification
7. Source the nominal full-scale resistance values for the 100-10M ranges summarized in
the Resistance verification table. Recalculate the limits based on the actual value of the
resistor and verify the reading is within the calculated limits.

Figure 10-7: Verifying four-wire resistance

Four-wire resistance verification data
Use the following values to verify the performance of the Series 3700. Actual values are
dependent on published specifications (see
3)).
Applied Resistance
Verify Res 100 Ohm
Verify Res 1k Ohm
Verify Res 10k Ohm
Verify Res 100k Ohm
Verify Res 1M Ohm
Verify Res 10M Ohm

Verifying two-wire resistance

Check the normal resistance function by connecting accurate resistance values to the Series
3700 analog backplane connector and verify that the displayed readings fall within specified
limits.
CAUTION
10-16
Range
1.00E+02
1.00E+03
1.00E+04
1.00E+05
1.00E+06
1.00E+07
Do not exceed 1000V peak between INPUT HI and INPUT LO because
instrument damage may occur.
Series 3700 System Switch/Multimeter Reference Manual
Calculating resistance reading limits
Value
Lower limit
1.00E+02
9.999310E+01
1.00E+03
9.999360E+02
1.00E+04
9.999360E+03
1.00E+05
9.999360E+04
1.00E+06
9.999560E+05
1.00E+07
9.995900E+06
Document Number: 3700S-901-01 Rev. A / August 2007
(on page 10-
Upper limit
1.000069E+02
1.000064E+03
1.000064E+04
1.000064E+05
1.000044E+06
1.000410E+07

Advertisement

Table of Contents
loading

Table of Contents