Section 11: Calibration
3. Send the following command:
SEND dmm.calibration.ac(2) print("done")
AC Cal Step 3: 100mV @ 50kHz
1. Source 100mV @ 50kHz
2. Send the following command:
SEND dmm.calibration.ac(3) print("done")
AC Cal Step 4: 1V @ 1kHz
1. Send the following command:
SEND dmm.range = 1
2. Source 1V @ 1kHz
3. Send the following command:
SEND dmm.calibration.ac(4) print("done")
AC Cal Step 5: 1V @ 50kHz
1. Source 1V @ 50kHz
2. Send the following command:
SEND dmm.calibration.ac(5) print("done")
AC Cal Step 6: 10V @ 1kHz
1. Send the following command:
SEND dmm.range = 10
2. Source 10V @ 1kHz
3. Send the following command:
SEND dmm.calibration.ac(6) print("done")
AC Cal Step 7: 10V @ 50kHz
1. Source 10V @ 50kHz
11-12
Series 3700 System Switch/Multimeter Reference Manual
Document Number: 3700S-901-01 Rev. A / August 2007