Keithley 3700 series Reference Manual page 263

System switch/multimeter
Hide thumbs Also See for 3700 series:
Table of Contents

Advertisement

Series 3700 System Switch/Multimeter Reference Manual
dmm.thermistor
Remarks
Example
dmm.thermocouple
Attribute
Usage
Remarks
Example
Document Number: 3700S-901-01 Rev. A / August 2007
This attribute is only valid when dmm.func = "temperature". All other
configurations generate an error and return nil when queried.
Changing functions with dmm.func has no effect on this setting.
The factory default and dmm.reset function value is 5000 ohms.
To set thermistor type to 3000:
dmm.func = "temperature"
dmm.thermistor = 3000
Indicates the thermocouple type.
To read the thermocouple type:
value = dmm.thermocouple
value: Represents the present thermocouple type
To write the thermocouple type:
dmm. thermocouple = value
value: Represents the desired thermocouple type. For value, use one of the following:
dmm.THERMOCOUPLE_J or 0
dmm.THERMOCOUPLE_K or 1
dmm.THERMOCOUPLE_T or 2
dmm.THERMOCOUPLE_E or 3
dmm.THERMOCOUPLE_R or 4
dmm.THERMOCOUPLE_S or 5
dmm.THERMOCOUPLE_B or 6
dmm.THERMOCOUPLE_N or 7
This attribute is only valid when dmm.func = "temperature". All other
configurations generate an error and return nil when queried.
Changing functions with dmm.func has no effect on this setting.
The factory default and dmm.reset function value is dmm.THERMOCOUPLE_K.
To set thermocouple type to J:
dmm.func = "temperature"
dmm.thermocouple = dmm.THERMOCOUPLE_J
Section 9: Instrument Control Library (ICL)
9-119