Section 10: Verification
3. Select the SLOW integration rate with the RATE key.
4. Set the Series 3700 for the 1 range, and make sure the FILTER is on. Enable OC+ (offset-
compensated ohms). (Use OC+ for 1 and 10range verification.)
5. Recalculate reading limits based on actual discrete resistor resistance values.
6. Repeat using the 10discrete resistor on the 10 range.
Figure 10-10: Verifying discrete resistance
Discrete resistance verification data
Use the following values to verify the performance of the Series 3700. Actual values are
dependent on published specifications (see
3)).
Applied Resistance
Verify Res 1 Ohm
Verify Dry Circuit 1 Ohm
Verify Res 10 Ohm
Verify Dry Circuit 10 Ohm
10-20
Series 3700 System Switch/Multimeter Reference Manual
Calculating resistance reading limits
Range
Value
1.00E+00
1.00E+00
1.00E+00
1.00E+00
1.00E+01
1.00E+01
1.00E+01
1.00E+01
Document Number: 3700S-901-01 Rev. A / August 2007
Lower limit
Upper limit
9.998600E-01
1.000140E+00
9.998500E-01
1.000150E+00
9.999310E+00
1.000069E+01
9.998500E+00
1.000150E+01
(on page 10-