Series 3700 System Switch/Multimeter Reference Manual
Figure 8-10: Standard event register
Commands to program and read the register are summarized in the table below. Refer to the
status.standard ICL command for the attributes to control bits.
Commands
*ESR?
or
print(status.standard.event)
*ESE <mask>
or
status.standard.enable = <mask>
*ESE?
or
print(status.standard.enable)
Document Number: 3700S-901-01 Rev. A / August 2007
Description
Read standard event status register.
Program the event status enable
register: <mask> = 0 to 255
Refer to the status.standard ICL
command
Read event status enable register.
Section 8: Status Model