AC volts calibration

Make the connections as shown below, then perform the following calibration steps (AC Cal
Step 1 through Step 10):

Figure 11-5: ACV verification

AC Cal Step 1: 10mV @ 1kHz
1. Send the following commands:
SEND dmm.func = dmm.AC_VOLTS
SEND dmm.range = 10e-3
2. Source 10mV @ 1kHz
3. Send the following command:
SEND dmm.calibration.ac(1) print("done")
AC Cal Step 2: 100mV @ 1kHz
1. Send the following command:
SEND dmm.range = 100e-3
2. Source 100mV @ 1kHz
Document Number: 3700S-901-01 Rev. A / August 2007