Keithley 3700 series Reference Manual page 348

System switch/multimeter
Hide thumbs Also See for 3700 series:
Table of Contents

Advertisement

Section 9: Instrument Control Library (ICL)
status.system2.*
Also see
Example
status.system3.*
Attribute
Usage
Remarks
9-204
status.system.*
(on page 9-201)
status.system3.*
(on page 9-204)
status.system4.*
(on page 9-205)
status.system5.*
(on page 9-207)
Sets the extension bit of the system2 enable register:
status.system2.enable = status.system2.EXT
TSP-Link system3 data structure register set.
To read condition, enable and event registers:
enablereg = status.system3.condition
enablereg = status.system3.enable
enablereg = status.system3.event
To write to enable register:
status.system3.enable = enablereg
Set enablereg to one of the following values:
0 Clears all bits.
To set EXT bit (B0)
1 or status.system3.EXTENSION_BIT
1 or status.system3.EXT
To set a node bit (Bn); n = 29 to 42.
status.system3.NODEn
In an expanded system (TSP- Link), this attribute is used to read or write to the
system3 node registers.
Reading a system3 node register returns a numeric value whose binary equivalent
indicates which register bits are set. The bits of the system3 node register are
identified as follows:
B0 - EXT
B4 - Node 32
B1 - Node 29
B5 - Node 33
B2 - Node 30
B6 - Node 34
B3 - Node 31
B7 - Node 35
For example, assume value 9 is returned for the enable register. The binary
equivalent is 0000000000001001. This value indicates that bit 0 (EXT) and bit 3
(Node 31) are set.
Assigning a value to the status.system3.enable attribute sets the extension bit or a
node bit of the system3 node enable register.
Series 3700 System Switch/Multimeter Reference Manual
B8 - Node 36
B12 - Node 40
B9 - Node 37
B13 - Node 41
B10 - Node 38
B14 - Node 42
B11 - Node 39
B15 - Not used
Document Number: 3700S-901-01 Rev. A / August 2007
.condition
.enable
.event
.condition
.enable
.event