Changing Filter Settings - Keithley 3700 series Reference Manual

System switch/multimeter
Hide thumbs Also See for 3700 series:
Table of Contents

Advertisement

Series 3700 System Switch/Multimeter Reference Manual
dmm.filter.window
Remarks
Also see
Example

Changing filter settings

To change the digital filter settings remotely, use the following commands:
To change the filter settings over the front panel, for the selected function:
1. Bring up the function attribute menu by pressing the CONFIG key then the DMM key
2. Turn the wheel to scroll and select the FILTER option.
3. Change the settings for enable, count, type and window.
Document Number: 3700S-901-01 Rev. A / August 2007
This is the filter window setting for the DMM and it applies to the selected function as
indicated by
dmm.func
does not support it will cause nil to be returned.
This attribute indicates the filter window to use in percent of range.
The command will generate an error when dmm.func = ―frequency‖, ―period‖, ―continuity‖
or ―nofunction‖. Also, an error will be generated if the value is out of range.
Changing functions with
The
dmm.reset
dmm.filter.enable
page 9-87).
To set the filter window for two-wire ohms to 0.25:
dmm.func = "twowireohms"
dmm.filter.window = 0.25
dmm.filter.count
(on page 9-86)
dmm.filter.enable
(on page 9-87)
dmm.filter.type
(on page 9-87)
dmm.filter.window
(on page 9-88)
(on page 9-91). Querying the setting when the selected function
dmm.func
(on page 9-91) has no effect on this setting.
(on page 9-112) function selects a filter window of 0.1.
(on page 9-87),
dmm.filter.count
Section 9: Instrument Control Library (ICL)
(on page 9-86),
dmm.filter.type
(on
9-89

Advertisement