Section 10: Verification
6. Verify the 10 range zero is within specification (see the 4-wire short applied resistance
verification table).

Figure 10-11: Four-wire short diagram

4-wire short applied verification data
Use the following values to verify the performance of the Series 3700. Actual values are
dependent on published specifications (see
3)).
Applied
Verify zeroes 100mV DC
Verify zeroes 1V DC
Verify zeroes 1 Ohm *
Verify zeroes 10 Ohm *
*The asterisk (*) designates that offset compensation (OC+) should be used on the range.
10-22
Series 3700 System Switch/Multimeter Reference Manual
Calculating resistance reading limits
Range
Current
1.00E-01
0.00E+00
1.00E+00
0.00E+00
1.00E+00
0.00E+00
1.00E+01
0.00E+00
Document Number: 3700S-901-01 Rev. A / August 2007
Lower limit
Upper limit
-9.000000E-07
9.000000E-07
-2.000000E-06
2.000000E-06
-8.000000E-05
8.000000E-05
-9.000000E-05
9.000000E-05