2. Send the following commands:
SEND dmm.calibration.dc(4,-10) print("done")
SEND print(errorqueue.count)
DC Cal Step 5: 100 Volt
1. Send the following command:
SEND dmm.range = 100
2. Source 100V.
3. Send the following commands:
SEND dmm.calibration.dc(5,100) print("done")
SEND print(errorqueue.count)

Resistance calibration

Perform the following calibration steps (DC Cal Step 6 through Step 9):

Figure 11-3: Verifying four-wire resistance

DC Cal Step 6: 100 Ohm
1. Send the following commands:
SEND dmm.func = dmm.FOUR_WIRE_OHMS
SEND dmm.range = 100
2. Source 100 Ohms then read the resistor value from the calibrator.
Document Number: 3700S-901-01 Rev. A / August 2007
Section 11: Calibration
11-7