Patient Data And Operating Status - Nellcor N-395 Operator's Manual

Table of Contents

Advertisement

Data Port Protocol
Patient Data and Operating Status
Time
TIME
%Sip02
BPM
PA
Status
12-NOV-99 14:00:05
400
f
20
50
The Time colurmi represents the N-395 real-time clock.
Patient Data
Patient data is highlighted in the display above. Parameter values
are displayed directly beneath the heading for each parameter. In
this example the %SpOz is 100, and the pulse rate is 190 beats
per minute. The "'*'next to the 190 indicates that 190 beats per
minute is outside of the alarm 1imits;indicated in the top row, for
pulse rate. If no data for a parameter is available three dashes
(-
-
-)
will be displayed.
PA is an indication of pulse amplitude. The number can range
from 0 to 254. There are no alarm parameters for this value. It
can be used for trending infoilnation as an indication of a change
in pulse volume, relative pulse strength, or circulation.
Operating Status
The Status colu~nn indicates alarm conditions and operating
status of the N-395. In this example, the PH means that the pulse
rate upper alarm limit (Pulse High) has been exceeded. A
complete listing of the status codes is listed in Table 1
1.
As
many as 4 codes can be displayed at one time in the Status
colunlrl.