Enclosure Leakage Current; Patient Applied Risk Current; Patient Isolation Risk Current - Nellcor OXIMAX N-595 Service Manual

Hide thumbs Also See for OXIMAX N-595:
Table of Contents

Advertisement

Enclosure
Leakage Current
Patient Applied
Risk Current
Patient Isolation
Risk Current
N-595
AC Line
Neutral Line
Cord
Cord
Closed
Closed
Closed
Closed
Closed
Open
Open
Closed
Open
Open
Open
Closed
AC Line
Neutral Line
Polarity
Normal
Closed
Normal
Open
Normal
Closed
Reversed
Closed
Reversed
Open
Reversed
Closed
AC Line
Neutral Line
Polarity
Normal
Closed
Reversed
Closed
Power Line
Ground
IEC 60601-1
Cable
Closed
100 µA
500 µA
Open
Closed
500 µA
500 µA
Closed
Closed
500 µA
500 µA
Open
Power Line
Ground
IEC 60601-1
Cable
100 µA
Closed
Closed
500 µA
500 µA
Open
Closed
100 µA
500 µA
Closed
OPen
500 µA
Power Line
Ground
IEC 60601-1
Cable
Closed
5 mA
Closed
5 mA
Specifications
AAMI/ANSI
Standard
ES1
100 µA
300 µA
300 µA
100 µA
300 µA
300 µA
AAMI/ANSI
Standard
ES1
10 µA
50 µA
50 µA
10 µA
50 µA
50 µA
AAMI/ANSI
Standard
ES1
50 µA
50 µA