Spo 2 Printed Circuit Board; Patient Interface; Power Supply; Power Monitoring - Nellcor OxiMax N-65 Service Manual

Handheld
Hide thumbs Also See for OxiMax N-65:
Table of Contents

Advertisement

SpO
Printed Circuit Board
2

User Interface Printed Circuit Board

O
M
N-65 Service Manual
XI
AX
The SpO
PCB provides patient interface and monitor operating power.
2

Patient Interface

The patient interface receives signals from the O
converted and supplied to the user interface PCB central processing unit (CPU).
The patient interface receives control signals from the CPU. These control signals
are used to control the light emitting diodes in the O

Power Supply

The power supply provides operating voltages to the SpO
interface PCB. These voltages are supplied to the:
SpO
PCB power monitoring function
2
SpO
PCB circuits
2
user interface PCB power supply
user interface PCB CPU
user interface PCB audio circuits

Power Monitoring

The SpO
power monitoring function monitors and controls the outputs of the
2
SpO
power supply. The SpO
2
interface CPU and SpO
2
supply output signals.
The user interface PCB provides:
User setup of the monitor via the keypad
Patient %SpO
and pulse rate via the display
2
Monitor status via the display and audio function
Printed reports via the IrDA (Infrared Data Association) and external
printer
Display readability in dark environments via the display backlight
Operating voltages for the user interface circuits
Operating voltage monitoring
XI
power monitoring function provides the user
2
patient interface with status information on the power
Technical Discussion
M
sensor. These signal are
AX
M
sensor.
XI
AX
PCB and the user
2
85

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents