Replacing Capillaries On A Semi-Micro And Micro Flow Cell - Agilent Technologies 1260 User Manual

Infinity diode array and multiple wavelength detector
Hide thumbs Also See for 1260:
Table of Contents

Advertisement

10
Maintenance

Replacing Capillaries on a Semi-Micro and Micro Flow Cell

Replacing Capillaries on a Semi-Micro and Micro Flow Cell

When
If the capillary is blocked
Tools required
Two 1/4 inch wrenches for capillary connections
Wrench 4 mm for capillary connections
Screwdriver Pozi 1 PT3
Parts required
Description
For parts see
Preparations
Turn the lamp(s) off.
Remove the front cover.
Remove the flow cell,
All descriptions in this procedure are based on the default orientation of the cell (as it is
N O T E
manufactured). The heat exchanger/capillary and the cell body can be fixed mirror
symmetrically to have both capillaries routed to the bottom or to the top (depending on the
routing of the capillaries to the column).
The fittings at the flow cell body are special types for low dead volumes and not compatible
N O T E
with other fittings.
When retightening the fittings, make sure that they are carefully tightened (handtight plus
1/4 turn with a wrench). Otherwise damage of the flow cell body or blockage may result.
214
"Semi-Micro Flow Cell Parts"
"Exchanging a Flow Cell"
on page 236 or
"Micro Flow Cell"
on page 200.
Agilent 1260 Infinity DAD and MWD User Manual
on page 238.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents