Maintenance of High Pressure Flow Cell

When
If the flow cell needs repair due to leaks or contaminations (reduced light throughput)
Tools required
• Two 1/4 inch wrenches for capillary connections hexagonal key 4 mm
• Tooth picks
Parts required
Description
For parts see
Preparations
• Turn the flow off.
• Remove the front cover.
• Remove the flow cell, see
All descriptions in this procedure are based on the default orientation of the cell (as it is
N O T E
manufactured). The heat exchanger/capillary and the cell body can be fixed mirror
symmetrically to have both capillaries routed to the bottom or to the top (depending on the
routing of the capillaries to the column).
Agilent 1260 Infinity DAD and MWD User Manual
"High Pressure Flow Cell"
"Exchanging a Flow Cell"

Maintenance of High Pressure Flow Cell

on page 248
on page 200.
10
Maintenance
207