Agilent Technologies 1260 User Manual page 23

Infinity diode array and multiple wavelength detector
Hide thumbs Also See for 1260:
Table of Contents

Advertisement

Table 3
Pin
1
2
3
4
5
6
7
8
9
Special Interfaces
Some modules have module specific interfaces/connectors. They are described
in the module documentation.
Agilent 1260 Infinity DAD and MWD User Manual
Remote Signal Distribution
Signal
Description
DGND
Digital ground
PREPARE
(L) Request to prepare for analysis (for example, calibration, detector
lamp on). Receiver is any module performing pre-analysis activities.
START
(L) Request to start run / timetable. Receiver is any module
performing run-time controlled activities.
SHUT DOWN
(L) System has serious problem (for example, leak: stops pump).
Receiver is any module capable to reduce safety risk.
Not used
POWER ON
(H) All modules connected to system are switched on. Receiver is any
module relying on operation of others.
READY
(H) System is ready for next analysis. Receiver is any sequence
controller.
STOP
(L) Request to reach system ready state as soon as possible (for
example, stop run, abort or finish and stop injection). Receiver is any
module performing run-time controlled activities.
START REQUEST
(L) Request to start injection cycle (for example, by start key on any
module). Receiver is the autosampler.