Nano Flow Cell - Replacing or Cleaning

Nano Flow Cell - Replacing or Cleaning

When
If parts are contaminated or leaky.
Tools required
Screwdriver POZI 1 PT3
Two 1/4 inch wrenches for capillary connections
Parts required
Description
For parts identification refer to
Preparations
Turn the lamp(s) off.
Remove the front cover.
Remove the flow cell, see
For details refer to the technical note that comes with the nano-flow cell kit.
N O T E
The quartz block can be cleaned with alcohol. DO NOT touch the inlet and outlet windows
N O T E
at the quartz block.
218
"Nano Flow Cells"
on page 244 (80 nL and 500 nL).
"Exchanging a Flow Cell"
Agilent 1260 Infinity DAD and MWD User Manual
on page 200.