4
While releasing the holder with a screw driver, carefully
insert the holmium oxide filter.
Next Steps:
6
Perform a holmium oxide test, see
filter.
7
Insert the flow cell, see
8
Replace the front cover.
9
Turn on the flow.
Agilent 1260 Infinity DAD and MWD User Manual
"Holmium Oxide Test"
"Exchanging a Flow Cell"
Cleaning or Exchanging the Holmium Oxide Filter
5
Replace the flow cell cover and fix the six screws.
on page 174 to check the proper function of the holmium oxide
on page 200.
10
Maintenance
225